Bionet CardioCare 2000 is an economically priced 12 channel resting ECG that does not compromise on performance. Suitable for use in private practice, ER, or hospital use, the CardioCare 2000 is designed for maximum ease of use and convenience. Software upgradeable and networkable, the CardioCare 2000 is an excellent choice for your diagnostic needs. |
- Simultaneous 12 Channel Resting ECG
- ECG Interpretation Program
- A4 Full Size Paper
- Low Maintenance Cost
- Fax Transmission Function
- Software Upgradeable
- Options
- Rechargeable Battery (ni-mh)
- Bag
- PC software (Pending FDA Approval)
- Limb electrode
- Chest electrode
- RS-232C null cable
- LAN cable (Ethernet-RJ-45)
|
- Dimension (H x W x D): 305.5 mm x 296 mm x 92.5 mm
- Weight: Approximately 2.98 kg
- Recording Channel: 3,6,12 channel and 1 channel (60 sec)
- Sensitivity: 5, 10, 20 auto (I~aVF: 10, V1~V6:5) mm/mV
- Printing Speed: 12.5, 25, 50 mm/s
- Display: 2 x 16 character LCD display
- Filters
- AC (50/60 Hz, -20dB or better)
- Muscle (25~35Hz, -3dB or better)
- Base Line Drift (0.1Hz, -3dB or better)
- Low pass filter (off, 40Hz, 100Hz, 150Hz)
- LCD display LED indicator: Sensitivity, speed, filter status, HR, printing form, rhythm lead signal quality, power source, battery
- Keyboard: Membrane keyboard Alphanumeric and symbol available
- Patient Data: ID, name, age, sex, height, weight
- Basic Measurement: Heart rate, PR, QRS, QT/QTc, P-R-T axis
- Signal qualilty control: Disconnected lead detection
- Battery capacity: 1 hour of normal use (approximately 100 automatic ECG printouts)
- Communication: PC connection with RS-232 interface and LAN
- Safety: Class 1, Type BF
- Conformity: CE, CSA
- Recorder: Thermal print head, thermal roll paper
- Report papers
- Width (A4): 210 mm or 8.5in
- Length (A4): 300 mm or 11in
- Effective writing width: 204 mm
- Resolution
- Vertical: 8 dots/mm
- Horizontal: 16 dots/mm
- Electrical
- Internal noise: 20uV(p-p) max
- Input circuit: Floating input
- Isolated and defibrillation protected
- Input impedance: >10MQ
- Input range: > +/- 5mV
- Common Mode Rejection: > +100dB
- DC offset voltage: > +/- 300mV
- Time constant: 3.2 sec
- Patient leakage current: <10uA
- Power
- Power Supply: AC or built-in Battery (option)
- Power requirement: 100 - 240 Vac, 50/60Hz, 1.0 - 0.5A
- Power consumption: 60w max
- Environment
- Operating humidity: 30~85%
- Operating temperature: 10° ~40° C
- Atmospheric pressure: 70 ~ 106 KPa
|
| Comes with a 3 Year Warranty |
|
|
|