- Dimensions (H x W x D): 305.5 mm x 296 mm x 92.5 mm
- Weight: Approximately 3.5 kg
- Recording channel: 3, 6, 12 channel and 1 channel (60 sec)
- Sensitivity: 5, 10, 20, auto (I~aVF: 10, V1~V6: 5) mm/mV
- Printing speed: 12.5, 25, 50 mm/s
- Display: 320 x 240 character LCD display (3 leads trends)
- Monitor Display: ID, date, sensitivity, speed, filter, form, rhythm lead
- User Interface: Touch screen (alphanumeric and symbol available), keyboard, rotary push-knob (pop-up menu)
- Patient data: ID, name, age, sex, height, weight
- Basic measurement: Heart rate, PR, QRS, QT/QTc, P-R-T axis
- Signal quality control: Disconnect lead detection
- Power: Power supply: AC or built-in battery (option) 100-240 vac, 50/60Hz, 1.0-0.5 A, 60W max
- Battery (Ni-MH): 1 hour of normal use (approx. 100 ECG printouts)
- Communication: PC connection with RS-232 interface and LAN
- Safety: Class 1, Type BF
- Conformity: CE, CSA
- Filters
- AC (50/60Hz, - 20dB or better)
- Muscle (25~35Hz, -3dB or better)
- Base line drift (0.1Hz, -3dB or better)
- Low pass filter: off, 40Hz, 100Hz, 150 Hz
- Recorder: Thermal print head, thermal roll paper
- Report papers
- Width (A4): 210 mm or 8.5in
- Length (A4): 300mm or 11in
- Effective writing width: 204 mm
- Resolution
- Vertical: 8 dot/mm
- Horizontal: 16 dot/mm
- Electrical
- Internal noise: 20uV (p-p )max
- Input circuit: Floating input
- Isolated and defibrillation protected
- Input impedance: ≥10MO
- Input range: ≥±5mV
- Common Mode Rejection: ≥+100dB
- DC offset voltage: ≥±300mV
- Time constant: 3.2sec
- Patient leakage current: <10uA
- Frequency response 0.05~150Hz
- Environment
- Operating humidity: 30~85%
- Operating temperature: 10° ~ 40° C
- Atmospheric pressure: 70 ~ 106 KPa
|