More than 3 million times every single day critical patient care decisions are made based on blood pressure determinations taken using DINAMAP technology. Now this technology is available in a durable, easy-to-use unit that fits even the most sensitive of budgets.
Pro 400 includes:
DINAMAP BP
Turbo-Temp
Masimo SET or Nellcor SpO2
Pulse Rate
|
| Enhanced memory stores up to 100 determinations within a 24 hr period.
Flexible mounting options.
Trim knob for flexible display.
Intra-arterial reference standard for blood pressuring using the oscillometric technology.
Water ingress protection.
Integrated power supply.
3 year warranty – parts and labor.
Options include ability to create an Electronic Medical Record for the HIS/CIS. |
Dimensions:
- Height: 9.8 in (25.0 cm)
- Width: 9.8 in (24.8 cm)
- Depth: 6.9 in (17.5 cm)
Weight: Including Battery 7.8 lb (3.5 kg)
Mountings: Self-supporting on rubber feet or pole mountable
Portability: Carried by recessed handle or pole mounted
Classification Information:
- Mode of operation: continuous
- Degree of protection against harmful ingress of water: Drip-proof IPX1
Power Requirements
- AC input voltage: 100-240 VAC, 50 / 60 Hz (nominal)
- 90 ~ 253 VAC, 47 ~ 63 Hz (range), 50VA.
- Protection against electrical shock: Class 1
- DC input voltage: 24 VDC (nominal), 12-30 VDC, 36VA, supplied from a source conforming to IEC 601-1.
- AC input is protected by two internal fuses, replaceable by qualified service personnel only. DC input line is protected by an internal auto-resetting fuse.
- Battery: 12 volt, 2.3 amp-hours protected by internal auto-resetting fuse.
- Minimum operation time: 2 hrs (5 min cycle with adult cuff at 25 ∞C with power save mode enabled) from full charge.
- Time for full recharge: 1 hr 50 min from full discharge when the Monitor is switched off and 8 hrs when the Monitor is switched on.
Environmental Operating Temperature + 5 ∞C to + 40 ∞C (+ 41 ∞F to + 104 ∞F)
Operating Atmospheric Pressure 700 hPa to 1060 hPa
Storage Temperature ñ 20 ∞C to + 50 ∞C (ñ 4 ∞F to + 122 ∞F)
Storage/Transportation:
Atmospheric Pressure 500 hPa to 1060 hPa
Humidity Range 0% to 95% noncondensing
Radio Frequency Complies with IEC Publication 601-1-2 (April 1993) Medical Electrical Equipment
Electromagnetic:
- Compatibility Requirements and Tests and CISPR 11 (Group 1, Class A) for radiated and conducted emissions.
|
|
|
|