Wallach’s Sonicaid One Fetal Doppler. The performance and reliability of the Sonicaid One provide reassurance for clinician and mother. The unit is so small and lightweight that it’s easy to transport and store. It’s simple to use and it’s comfortable to hold. One model meets every requirement. Higher sensitivity with wider beam coverage enables location of heart signals quickly from early gestation to full term. Fetal heart sounds are heard in excellent clarity and the large, reflective display can be easily read. The recessed buttons and “memory” volume control add to the unique styling and work-friendly ergonomics to create the ultimate in ease of use. Advances in materials have enabled a lighter, yet more rugged unit including the curly “duraflex” probe cable for increased reliability and extended product life. Over 1000 one-minute examinations and auto-shutoff mean there is no fetal heart detector with better economy. |
- Unique design
- Easy to use
- Waterproof probe as standard
- Fetal heart rate display
|
- Physical
- Size Main unit: 128 mm x 68 mm x 38 mm
- Probe: 71 mm x 32 mm Ø
- Weight: 230 g
- Power Supply
- Battery: 2x AA 1.5V Alkaline or 2x AA 1.2VNiCad
- Battery life: 1000 one minute examinations (Alkaline)
- Auto Shut off: 5 min in use / 2 min no signal
- Unit Controls
- Ultrasound
- Frequency: 2MHz Continuous Wave
- Acoustic Output: Low Power - exemption BS EN 61157 (Clause 6)
- Outputs
- Loudspeaker: 350mW rms typical
- Headset: 10mW total (32Q stereo headset)
- Audio Response: 350Hz - 1 kHz
- Fetal Heart Rate
- Display: 3 digit FHR, Pulse/Signal quality, Battery Low
- Range: 50-210 bpm
- Accuracy: < ± 2 bpm over 100-180 bpm
- Resolution: ± 1 bpm
- Averaging: 3 - 5 beats
- Safety
- Classification Class 1 Type: EN60601-1, IEC60601-1, CAN/CSA-C22.2NO.601.1, UL2601-1 and EN60601-1-2
|
|
|
|